SA 2 Firmware Update

To keep your SA 2 system optimized to perform at its highest level, you must upgrade your system to the most recent firmware/software version.

The latest SA 2 update [Version 1.8.2.1] includes the addition of DICOM PACS uploading, a fix for spaces in the Session Name for workflows, improvements to doppler, and a completely redesigned video pipeline. Additionally there are R&D improvements and miscellaneous bug fixes. 

Download firmware update for SA 2 here: [Version 1.8.2.1]

  1. Load the upgrade file on a thumb drive [USB memory stick] and insert into the lower SA USB slot.
  2. Hold the power button while powering on your EVO portable ultrasound system—release the power button when the keyboard starts blinking.
  3. Allow the scanner to boot and follow the on-screen prompts.

Once the system has rebooted, we recommend you reset the SA to factory default settings. This will refresh all of the probe exam types to new settings to provide the best possible image quality.

  1. Press SUPER from the HOME menu.
  2. Press F2 [SETUP]
  3. Press F4 [SYSTEM]
  4. Press F2 [RESET]
  5. Choose RESET ALL option
  6. Confirm by selecting YES at the prompt

If you would like to receive further instructions or assistance, we have a customer service representative waiting to assist you during our business hours [see sidebar]. 